Skip to main content
StreamVetree
Sign in
OphthalmologyCase series / Retrospective2 min read · distilled by Vetree AI

Validation of the iFalcon V100 Rebound Tonometer in Canine Eyes.

Su Y, Liu Y, Shi F, Fu J · Veterinary Ophthalmology · 1 March 2026

Clinical bottom line

iFalcon V100 accurately measures canine IOP across physiological and mild pathological ranges.

Summary

This validation study assessed the iFalcon V100 rebound tonometer's accuracy for measuring intraocular pressure (IOP) in canine eyes using four freshly enucleated normal eyes. Researchers created a controlled pressure system using a water syringe in the vitreous chamber and an implantable pressure sensor in the anterior chamber, testing pressures from 5 to 50 mmHg. The iFalcon V100 demonstrated strong linear correlation with direct sensor readings (r = 0.96) and minimal bias (+0.34 mmHg), with 95% limits of agreement from -4.29 to +4.98 mmHg. The device showed slightly higher readings at lower pressures and required multiple attempts at 5-10 mmHg. The validated range of 5-50 mmHg encompasses normal physiological pressures and common pathological levels, though the study did not evaluate performance during acute glaucoma (70-80 mmHg). Overall, the iFalcon V100 provides clinically acceptable and accurate IOP measurements in normal canine eyes within this range.

OphthalmologySmall Animal

This summary was distilled by AI and may occasionally misinterpret data. Confirm critical details with the primary literature before clinical application.